Buscador de publicaciones<< Volver

Núñez-Gil IJ, Echavarría M, Escaned J, Biagioni C, Feltes G, Fernández-Ortiz A

Bioresorbable stent restenosis: new devices, novel situations.

J Invasive Cardiol. 2014 Dec;26(12):E164-6, PMID: 25480999

A 58-year-old man presented to our hospital with effort angina. Ten months prior, he was treated with a Bioresorbable vascular scaffold (BVS). During the current admission, an image angiographically compatible with in-BVS restenosis at the circumflex ostium with a radiolucent image in the ostial left anterior descending artery was shown. BVS failure is very infrequent and this is one of the first cases of BVS restenosis described. Thus, data on the best management option are scarce. We treated it like a drug-eluting stent restenosis, performing first an intracoronary optical coherence tomography scan in order to identify the left descending radiolucent image and to prepare the best treatment strategy.

CARDIAC & CARDIOVASCULAR SYSTEMS

Una iniciativa de

Ministerio de Economía y Competitividad Fondo Europeo de Desarrollo Regional IMIM - Parc de Salut Mar